Haefely Hipotronics AXOS 5 Compact Immunity Test System

Description

Haefely Hipotronics AXOS 5 Compact Immunity Test System integrates all the best features of a standalone test systems into one single economic solution. It is capable of 5kV EFT/Burst, 5kV Combination Wave (1.2/50us & 8/20us), AC/DC Dips & Interrupts and Pulsed Magnetic Field. The unit also includes an integrated single-phase coupling decoupling network (CDN) for EMC testing applications up to 16A.The expandable functionality allows for quick and completely automated testing to the most common IEC, EN, ANSI, IEEE and UL standards.

 

MODEL SERIES OVERVIEW

Haefely Hipotronics AXOS-series Expandable Conducted Immunity Test Systems

Features

  • Manual and automated test modes
  • Economic & Efficient Touch screen
  • Safety interlock, warning lamp and emergency stop functions
  • Voltage and current monitoring
  • Automatic generation of test report
Download Data Sheet
Print
Email

Specifications

GENERAL DATA
Control power 85 V – 264 V

50/60 Hz

User test storage unlimited
Remote interface Ethernet, RJ45
Display 7“/ 800×480 / 24 bit with touch-screen
USB for USB memory stick
AUX. interface D-sub 37p for external CDN, external transformer etc.
External trigger input 5 V TTL
Synch input BNC, 10 V – 264 V AC
Trigger output 5 V TTL
External start / stop input 5 V TTL, starts / stops predefined test sequence
EUT failed input 5 V TTL
Analog output 0 – 10 V, for use with external options
Warning lamp output 2 x 24 V / 1 A DC
Safety circuit stops the test when unlocked
Dimensions 19“/ 4U
(W x H x D)  (45 x 18 x 49 cm)
Weight 30 kg
IEC / EN 61000-4-4 EDITION 2 EFT / BURST
Output voltage 0.2 – 5.0 kV ±10% at coaxial output
Spike frequency 1 Hz – 1 MHz
Polarity pos / neg / alternate
Burst duration 10 us – 1 s
Output impedance 50 Ohms
Burst period 1 ms – 10 s
Rise time 5 ns ±30%
Test time 1 s– 999 minutes
Impulse duration 50 ns ±30% at 50 Ohm

50 ns –15 +100 ns at 1000 Ohm

Trigger automatic, manual,

external trigger input

Burst mode normal, continuous, real, random
Integrated single phase coupling / decoupling network 264 V AC / 16 A

220 V DC / 10 A

IEC / EN 61000-4-5 EDITION 2 SURGE COMBINATION WAVE
Output voltage 0.2 – 5.0 kV ±10% Output current 0.1 – 2.5 kA ±10%
Voltage rise time 1.2 us ±30% Current rise time 8 us ±20%
Voltage duration 50 us ±20% Current duration 20 us ±20%
Polarity pos / neg / alternate
Integrated single phase CDN 264 V AC / 16 A

220 V DC / 10 A

Output impedance 2 Ohms
Phase synch 0 – 359° with 1° steps or asynchronous mode
Impulse trigger automatic 5 s – 99999 s

manual external trigger input

Counter preselect counter 1 – 10000 / infinite 100000
Peak voltage monitor BNC output: 1000:1

display: 3 digits

Peak current monitor BNC output: 1 kA/V

display: 3 digits

IEC / EN 61000-4-11 EDITION 2 AND IEC / EN 61000-4-29 DIPS & INTERRUPTS
Max. voltage 264 V AC/DC Interrupt time 0.1 period – 9999 periods

0.02 ms – 999 minutes

Max. current 16 A AC/DC continuous

20 A for 5 s

40 A for 3 s

500 A inrush current

Interval time 1 period – 9999 periods synch

0.04 ms – 999 minutes synch

Trigger automatic

Manual external trigger input

Test time 1 s – 999 minutes

infinite

Interrupt dip level 0%

0% – 99% with external voltage source

Phase synch 0 – 359° 16 / 40 / 50 / 60 Hz

asynchronous mode

RMS voltage monitor BNC output: 100:1

display: 4 digits

RMS current monitor BNC output: 10 A/V

display: 4 digits

Stay informed