Haefely Hipotronics AXOS 8 Compact Immunity Test System
Description
Haefely Hipotronics AXOS 8 Compact Immunity Test System integrates all the best features of a standalone test systems into one single economic solution. It is capable of 5kV EFT/Burst, 7kV Combination Wave (1.2/50us & 8/20us), 7kV Ring Wave, AC/DC Dips & Interrupts, Pulsed Magnetic Field, and Telecom Wave. The unit also includes an integrated single-phase coupling decoupling network (CDN) for EMC testing applications up to 16A.The expandable functionality allows for quick and completely automated testing to the most common IEC, EN, ANSI, IEEE and UL standards.
MODEL SERIES OVERVIEW
Haefely Hipotronics AXOS-series Expandable Conducted Immunity Test Systems
Features
- Manual and automated test modes
- Economic & Efficient Touch screen
- Safety interlock, warning lamp and emergency stop functions
- Voltage and current monitoring
- Automatic generation of test report
Specifications
GENERAL DATA | |
Control power | 85 V – 264 V
50/60 Hz |
User test storage | unlimited |
Remote interface | Ethernet, RJ45 |
Display | 7“/ 800×480 / 24 bit with touch-screen |
USB | for USB memory stick |
AUX. interface | D-sub 37p for external CDN, external transformer etc. |
External trigger input | 5 V TTL |
Synch input | BNC, 10 V – 264 V AC |
Trigger output | 5 V TTL |
External start / stop input | 5 V TTL, starts / stops predefined test sequence |
EUT failed input | 5 V TTL |
Analog output | 0 – 10 V, for use with external options |
Warning lamp output | 2 x 24 V / 1 A DC |
Safety circuit | stops the test when unlocked |
Dimensions | 19“/ 4U |
(W x H x D) | 45 x 18 x 49 cm |
Weight | 30 kg |
IEC / EN 61000-4-4 EDITION 2 EFT / BURST | |
Output voltage | 0.2 – 5.0 kV ±10% at coaxial output |
Spike frequency | 1 Hz – 1 MHz |
Polarity | pos / neg / alternate |
Burst duration | 10 us – 1 s |
Output impedance | 50 Ohms |
Burst period | 1 ms – 10 s |
Rise time | 5 ns ±30% |
Test time | 1 s– 999 minutes |
Impulse duration | 50 ns ±30% at 50 Ohm
50 ns –15 +100 ns at 1000 Ohm |
Trigger | automatic, manual,
external trigger input |
Burst mode | normal, continuous, real, random |
Integrated single phase coupling / decoupling network | 264 V AC / 16 A
220 V DC / 10 A |
IEC / EN 61000-4-5 EDITION 2 & 3 SURGE COMBINATION WAVE | |
Output voltage | 0.2 – 7.0 kV ±10% Output current 0.1 – 3.5 kA ±10% |
Voltage rise time | 1.2 μs ±30% Current rise time 8 us ±20% |
Voltage duration | 50 μs ±20% Current duration 20 us ±20% |
Polarity | pos / neg / alternate |
Integrated single phase CDN | 264 V AC / 16 A
220 V DC / 10 A |
Output impedance | 2 Ohms |
Phase synch | 0 – 359° with 1° steps or asynchronous mode |
Impulse trigger | automatic 2 s – 100 min
manual external trigger input |
Counter preselect counter | 1 – 1000 / infinite |
Counter | 100000 |
Peak voltage monitor | BNC output: 1000:1
display: 3 digits |
Peak current monitor | BNC output: 1 kA/V
display: 3 digits |
IEC / EN 61000-4-11 EDITION 2 AND IEC / EN 61000-4-29 DIPS & INTERRUPTS | |
Max. voltage | 264 V AC/DC |
Interrupt time
|
0.5 period – 800 periods
100 μs – 1000 minutes |
Max. current | 16 A AC/DC continuous
20 A for 5 s 23 A for 3 s 40 A for 3 s > 500 A inrush current |
Interval time | 1 period – 800 periods synch
100 μs – 1000 minutes asynch |
Trigger | Automatic
Manual external trigger input |
Test time | 1 s – 1000 minutes infinite |
Interrupt dip level | 0%
0% – 99% with external voltage source |
Phase synch | 0 – 359° 16 / 40 / 50 / 60 Hz
asynchronous mode |
RMS voltage monitor | BNC output: 100:1
display: 4 digits |
RMS current monitor | BNC output: 10 A/V
display: 4 digits |
IEC / EN 61000-4-12 EDITION 2 AND ANSI / IEEE C62.41 Ring Wave | |
Max. voltage | 0.2 – 7.0 kV ± 10% |
Repetition rate | Up to 30 pulses / min |
Frequency | 100 kHz |
Polarity | Positive / negative / alternate |
Rise time OC | 5 μs |
Rise time SC | 1 μs |
Floating output | Max. 460 V / AC |
Phase sync accuracy | ± 1° |
Impedance | 12 Ohm, 30 Ohm |
Damping rate | 0.4 < peak1/peak2 < 1.1
0.4 < peak3/peak2 < 1.1 0.4 < peak4/peak3 < 1.1 |
Peak voltage monitor | BNC output: 1000:1
Display: 3 digits |
Peak current monitor | BNC output: 1kA / V
Display: 3 digits |
- ANSI/IEEE Standards
- ANSI/IEEE C62.41
- Basic Standards
- IEC 61000-4-10
- IEC 61000-4-11
- IEC 61000-4-12
- IEC 61000-4-29
- IEC 61000-4-4
- IEC 61000-4-5
- IEC 61000-4-8
- IEC 61000-4-9
- Telecom Standards
- ITU-T K.12
- ITU-T K.20
- ITU-T K.21
- ITU-T K.28
- ITU-T K.41
- ITU-T K.44
- ITU-T K.45
- Product Standards
- EN 300 386 V1.3.2
- EN 50370-1
- EN 61643
- FCC 97-270 (Part 68)
- IEC 60255-1
- IEC 60601-1-2
- IEC 61009-1
- IEC 61326-1
- IEC 61850-3
- EN 301 489-1
- EN 301 489-17
- EN 301 489-24
- EN 301 489-7
- Generic Standards
- EN 61000-6-1
- EN 61000-6-2